Call up the PC 600 Carrier memory file.
Using 100 mL of CVS blank solution under the electrode, run conditioning cycles as
needed, then run the calibration using the 3% Carrier Standard for addition.
Run 3% Carrier Standard as sample by using 100 mL of fresh CVS blank solution under
the electrode and 3% Carrier Standard for addition. The result should be between 2.8%-
3.2%, otherwise repeat conditioning and calibration.
Run the sample by using 100 mL of fresh CVS blank solution under th